Range: 0.5’
to 33’
Accuracy: ±0.25%
of span (air)
Resolution: 0.125
(3 mm)
Beam width: 8°
conical
Dead band: 0.5’
Air Temperature: F: -40º to 140º
C:
-40º to 60º
Temp. comp: Automatic
Pressure rating: 30 psi (2 bar) @ 25°C.,
derated
@ 1.667 psi (.113
bar) per °C. above 15
°C.
Rec. cable: (5)
conductor, shielded
Cable length: 150’
(48m) max.
Enclosure: NEMA
4X (IP 65)
Installed height: 4.1” (10.4 cm)
Encl. material: PP UL94VO
Trans. material: PVDF Kynar®
Mount. threads: 2” NPT (2”
G)
Mount. gasket: Viton®(G version
only)
Conduit entrance: Single, 1/2” NPT/M25
Controller:
Display type: 2-line
LCD, 16 digit
Display units: Distance:
inch/cm
Volume: gallon/liter
Memory: Non-volatile
Supply voltage: 120/240
VAC @
50-60
Hz.
Signal output: 4-20
mA two-wire
Current flow: Isolated,
sinking
Signal invert: 4-20
or 20-4 mA
Calibration: Digital,
push button
Contact type: (5)
SPDT
Contact rating: 250
VAC @ 10A
Relay fail-safety: Reverts to de-energized
state
during input-loss
Relay indication: LED, relay status
Temperature: F:
-40º to 140º
C:
-40º to 60º
Enclosure: NEMA
4X (IP65)
Encl. material: Polycarbonate
Conduit entrance: 1/2” Knock outs
Classification: General
purpose
CE compliance: EN 50082-2
EN
55011
EN
61010-1 |